- DC voltage
- Measurement method: Keysight patented continuously integrating multi-slope IV A/D converter
- A/D Linearity: 0.0001% of reading + 0.0001% of range
- Input resistance
- 0.1 V, 1V, 10V range: Selectable 10 MΩ or >10 GΩ
- 100 V, 300 V range: 10 MΩ ± 1%
- Input bias current: < 30 pA at 25 ºC
- Input protection: 300 V on all ranges
- True RMS AC voltage
- Measurement type: AC-coupled True RMS. Measures the AC component of the input.
- Measurement method: Digital sampling with anti-alias filter
- Maximum input: 250 DCV, 300 Vrms
- Input impedance: 1 MΩ ± 1%, in parallel with < 100 pF
- Input protection: 300 Vrms
- DC current
- Input protection: 1.6A 250V fuse on DAQM901A
- True RMS AC current
- AC measurement method: Digital sampling with anti-alias fil
- Power consumption: 45 watts