- Range: –80 °C to 110 °C
- Stability:
- ±0.006°C at –80°C (ethanol)
- ±0.005°C at 0°C (ethanol)
- ±0.005°C at 100°C (oil 5012)
- Uniformity:
- ±0.007 °C at –80 °C (ethanol)
- ±0.007 °C at 0 °C (ethanol)
- ±0.007 °C at 100 °C (oil 5012)
- Heating Time: 60 minutes, from 25 °C to 150 °C (oil 5012)
- Cooling Time: 210 minutes, from 25 °C to –20 °C (ethanol)
- Stabilization Time: 15–20 minutes
- Temperature Setting: Digital display with push-button data entry
- Set-Point Resolution: 0.01°; 0.00018° in high-resolution mode
- Display Resolution: 0.01 °
- Digital Setting Accuracy: ±1°C
- Digital Setting Repeatability: ± 0.01 °C
- Access Opening: 120 x 172 mm (4.7 x 6.8 in)
- Depth:
- 457 mm (18 in) without Liquid-in-Glass Thermometer Cal Kit
- 482 mm (19 in) with Liquid-in-Glass Thermometer Cal Kit
- Wetted Parts: 304 stainless steel
- Power: 230 VAC (±10 %), 50 or 60 Hz, specify, 10 A
- Volume: 15,9 Liters