- Measurement method:
- Conductivity: AC 2-Electrode Method
- Temperature: Thermistor Resistor
- Conductivity Measuring frequency: 80Hz/3kHz Auto-selection
- Display unit: Customized LCD
- Measurement Item/Range:
- Conductivity: Depends on the cell
- Resistivity: Depends on the cell
- Salinity: Conversion from conductivity
- TDS: Conversion from conductivity
- Temperature: 0.0 ~100.0℃
- Display Range:
- Conductivity (Manual/Auto Range Switching) (Depending on the cell):
- 0.0 ~ 200.0μS/m (0.000 ~ 2.000μS/cm)
- 0.000 ~ 2.000mS/m (0.00 ~ 20.00μS/cm)
- 0.00 ~ 20.00mS/m (0.0 ~ 200.0μS/cm)
- 0.0 ~ 200.0mS/m (0.000 ~ 2.000mS/cm)
- 0.000 ~ 2.000S/m (0.00 ~ 20.00mS/cm)
- 0.00 ~ 20.00S/m (0.0 ~ 200.0mS/cm)
- 0.0 ~ 200.0S/m (0.000 ~ 2.000S/cm)
- SI unit (S/m) or (S/cm) selectable
- Resistivity (Manual/Auto Range Switching) (Depending on the cell):
- 0.005 ~ 2.000 Ω・m (0.5 ~ 200.0Ω・cm)
- 0.00 ~ 20.00 Ω・m (0.000 ~ 2.000kΩ・cm)
- 00.0 ~ 200.0 Ω・m (0.00 ~ 20.00kΩ・cm)
- 0.000 ~ 2.000kΩ・m (0.0 ~ 200.0kΩ・cm)
- 0.00 ~ 20.00kΩ・m (0.000 ~ 2.000MΩ・cm)
- 00.0 ~ 200.0kΩ・m (0.00 ~ 20.00MΩ・cm)
- 0.000 ~ 2.000MΩ・m (0.0 ~ 200.0MΩ・cm)
- 0.00 ~ 20.00MΩ・m (0 ~ 2000MΩ・cm)
- SI unit (Ω・m) and conventional unit (Ω・cm) selectable
- Salinity:
- 0.00 ~ 4.04% (NaCl)
- 0.00 ~ 42.40 (PSS:Sea water salinity)
- TDS (Manual/Auto Range Switching):
- 0 ~ 99.99mg/L
- 0 ~ 999.9mg/L
- 0 ~ 9.999g/L
- 0 ~ 99.99g/L
- 0 ~ 999.9g/L
- Temperature: -5.0 ~110.0℃
- Repeatability(Meter main unit):
- Conductivity: ±0.5% FS
- Resistivity: ±0.5% FS
- Salinity: ±0.5% FS
- TDS: ±0.5% FS
- Temperature: within ±0.2℃
- Temperature Compensation:
- Temperature Compensation Range:
- ATC (Auto Temperature Compensation):0 ~100.0℃
- MTC(Manual Temperature Compensation):0 ~100.0℃
- N/A: ATC OFF
- Standard Temperature Setting: 25℃ fixed
- Temperature Coefficient(Linear): 0 ~10.00%/℃
- Temperature Calibration: 1-point Calibration
- Performance Guaranteed Temperature, Humidity: 0 ~ 45℃ 85% or below(condensation-free)
- Data Memory: 1000 data
- Print Function: Connectable to optional external printer EPS-P30
- Auto Hold Fuction: yes
- Calibration History Creation Function: Latest one run
- Interval Measurement: yes
- External Input/Output:
- RS-232C Interface: yes (External Printer)
- USB (peripheral, Micro): yes
- Option Connection:
- External Printer: yes
- Power Source: AA Alkaline battery×4 or USB power feeding (No charging function).
- Power Consumption: Approx.1000 hours (estimated)
Product sets
- Main unit
- Sensor Holder: 7430850K
- Sensor stand (Stand, strut, stopper): 7430860K
- Tilt stand: 7430870K
- Sensor attachment (G): 0IB00004
- Conductivity sensor: CT-58101B
- Ground wire (2m): X0979500
- Beakers (150mL): 0DE00001
- AA Battery (4pcs)
- Manual
Reviews